Instruction manual

CHAPTER 2 : OPERATION GUIDE
86
List Window
Operation
Linked target data
The linked target data is displayed when List - Categorized List - Show Linked Target is checked in
the Data List tab of the Display Settings dialog box. To show this dialog box, select Tool - Vi e w S et-
tings from the menu bar. The lines of the linked target data cannot be scrolled.
Statistical value
The statistical values of the sample data linked with the target data are displayed.
The statistics are displayed when List - Categorized List - Show Statistics is checked in the Data List
tab of the Display Settings dialog box. To show this dialog box, select Tool - View Settings from the
menu bar. The lines of the statistics cannot be scrolled.
When Multi-Point measurement is also specified, the row of Non-Uniformity (color variations) is
displayed.
The averaging calculation performed here averages the colorimetric data of each piece of data that
was calculated individually according to its spectral data or XYZ data.
On the other hand, the averaged measurement (described on page 50) and the averaging of list data
(on page 90) first average spectral data or XYZ data to obtain data that is then used in the calculation
of colorimetric data. For this reason, the results of these two types of calculations may differ.
Display example for the Light-source Color - RGB & Contrast Mode measurement
Linked target data
Statistical value
Linked target data
Statistical value
Display example for the Light-source Color - Normal Mode, Multi-Point measurement
(No. of points: 5)