Specications
Type Chlorophyll Meter SPAD-502Plus
Measurement subject Crop leaves
Measurement method Optical density dierence at 2 wavelengths
Measurement area 2 mm×3 mm
Subject thickness 1.2 mm maximum
Subject insertion depth 12 mm (with stopper having position adjustable from 0 to 6 mm)
Light source 2 LED elements
Receptor 1 SPD (silicon photodiode)
Display
LCD panel showing 4-digit measurement value (values shown to rst decimal place) and 2-digit number of
measurements; Trend graph of values in memory can also be shown.
Display range -9.9 to 199.9 SPAD units
Memory function Memory capacity for up to 30 values; Calculation/display of average of data in memory also possible
Power source 2 AA-size alkaline batteries
Battery life More than 20,000 measurements (when using new alkaline batteries under Konica Minolta test conditions)
Minimum interval between measurements Approx. 2 seconds
Accuracy
Within ±1.0 SPAD units (for SPAD value between 0.0 and 50.0 under normal temperature/humidity)
“*” added to display when measurement exceeds 50.0 SPAD units
Repeatability Within ±0.3 SPAD units
For SPAD value between 0.0 and 50.0 (with no change in sample position)
Reproducibility Within ±0.5 SPAD units
Temperature drift Within ±0.04 SPAD units/°C
Operating temperature/humidity range 0 to 50°C; Relative humidity of 85% or less (at 35°C) with no condensation
Storage temperature/humidity range -20 to 55°C; Relative humidity of 85% or less (at 35°C) with no condensation
Size; Weight 78 (W) ×164 (H) ×49 (D) mm, 200 g (excluding batteries)
Other Warning buzzer; User compensation factor
Standard accessories Depth stop; Strap; 2 AA-size alkaline batteries; Soft case; Reading checker
SPAD value: Index value displayed by Konica Minolta Chlorophyll meters and having a correlation to chlorophyll density.
Specications and appearance subject to change without notice.
164
7849
(Units: mm)
9242-4856-10 AJKAPK Printed in Japan1©2009 KONICA MINOLTA SENSING,INC.
Dimensions
SAFETY PRECAUTIONS
For correct use and for your safety, be sure to read the
instruction manual before using the instrument.
Be sure to use the specified battery. Using improper
battery may cause a re or electric shock.
Registration No. : YKA 0937154
Registration date : March 3, 1995
Registration No. : JQA-E-80027
Registration date : March 12, 1997