User Manual

SpirolabIII – User’s Manual Code MIR 980067 REV 0 Page 30 / 47
If it is not possible to make the calculations for lack of data, the interpretation is not valid and this message appears:
Not Valid
For more information on the methods of test interpretation see the attached flow chart.



 >'  >
2.10 OXIMETRY TESTING
SpirolabIII can carry out different types of oximetry tests, which are described in the following paragraphs.



 SpirolabIII    '   )            ) 
'  
)  3 2 ' '  
SpirolabIII' M"' '  
' .



' SpirolabIII
  =  .2,         D  

0  ) SpirolabIII' ' 
 ' '   
For the non-invasive measurement of the SpO
2
oxygen saturation and blood pulse rate utilize the re-usable finger
sensor. This sensor is recommended for patients weighing > 20 Kg with limited activity.
SpirolabIIImemorises the two oximetry values every 2 seconds.
Carry out an oximetry test as follows:
Connect the sensor to the instrument: insert the connector with the
arrow (printed on the connector) face-up, as shown:
Choose a high perfusion site, which is easily adaptable to the
sensor.
Insert finger into the sensor until the finger touches the end of the
probe. Ensure that the bottom part of the finger completely covers
the detector. If the finger is not able to be correctly positioned, use
another finger.
Position the sensor so that the cable is underneath the palm of the
hand. This enables the light source to remain on the fingernail and
the detector on the bottom part of the finger. From the main screen
press to access the test menu.
Select required test using the vertical scroll arrow.
If this message appears:
OXIMETRY NOT DETECTED
then your instrument does not include an oximeter.
If this message appears:
OXIMETRY DISABLED
the function has not been enabled. In this case contact a service centre or the manufacturer.
The oximetry tests that can be performed by SpirolabIII are:
Walk test - 6MWT
SLEEP oximetry
Oximetry (SpO2/BPM)