User Manual

spirobank G User Manual cod. 980026 Rev 2.0 Page 44 of 46
Guidance and manufacturer’s declaration – electromagnetic immunity
The spirobank G is intended for use in the electromagnetic environment specified below. The customer
or the user of the spirobank G should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±8 kV air
±
6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
±
1 kV differential
mode
±2 kV common mode
Not Applicable Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips,
short
interruptions and
voltage
variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Not Applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
[3 ] V
[3 ] V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the spirobank G, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d=[ 3.5
] P
3
d=[ 3.5
] P 80 MHz to 800 GHz
3
d=[ 7
] P 800 MHz to 2,5 GHz
3
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation