User Manual

Spirodoc cod. 980156 Rev 1.3 EN 29/67
ENGLISH
devices are found the display will list these devices with their respective names. By
touching the device of interest it can be memorized as a printer, a phone, or as a PC
On line; select an option.
In the “Bluetooth settingmenu the memorized devices can be viewed in the “printer”
list, “telephone” list and the “PC On line” list. Any device from these lists can be set
as the default device (the device that SPIRODOC will automatically connect to via
Bluetooth) by entering in the lists, touching the display and selecting the device. A listed
device can be eliminated from the list. (in this specific case the user will confirm the
deletion with the OK icon.)
So as not to make any modifications touch the icon in the bottom left side. In
the “Bluetooth setting” menu the “Bluetooth ON/OFF” option allows to enable/disable
the Bluetooth function thereby improving the battery consumption of the device.
Firmware Information
The user can access the menu to view information of the version control of the
following components:
Spirodoc
Bluetooth
After approximately 10 seconds the device automatically displays the Service Menu,
otherwise press .
Oximetry Alarms
Access to this function allows the
setup of reference values for SpO2
and BPM; an acoustic alarm will warn
the user if the SpO2 and BPM during
a test fall below the minimum or rise
above the maximum set reference
SpO2 and BPM values.
The cursor “_” in the beginning will be set on the minimum SpO2 value, use the
visualized numbers to select the desired value. Move on to the next value by touching
OK; once all four values have been set the user may activate or deactivate the acoustic
alarm: Touch ON to activate the acoustic alarm during oximetry testing, or touch OFF to
deactivate the acoustic alarm. To confirm the desired selection touch OK in the bottom
right side of the display, automatically the device will return to the service menu.
SpO2 sampling rate
This function allows to set the time that elapses between values memorised between
consecutive oximetry parameters; touch one of the two visualized icons: 2 seconds or 4
seconds, then touch OK to set the selected value and the device will automatically
return to the service menu.
Spirodoc mode
The device can be setup in one of the following modes: