User's Manual

Spirobank II
Page 14 of 34
Rev.1.1
User manual
The specifications for both the oximetry and for the cardiac pulse are the same regardless of which of the above mentioned oximetry
sensors is used.
1.6.3 Other features
Memory
Memory capacity for over 10000 spirometric tests
The precise number depends on the individual configuration, so it cannot be
determined more closely
keyboard
membrane keyboard with 6 keys
Display
Display LCD 160x80 monochromatic
Interface
USB, Bluetooth
Bluetooth interface
frequenzy range = 2402-2480 MHz
rated RF power output = 7.5 dBm maximum trasmit power
type of antenna = drawn on the board
Duration of the 3,7V lithium battery
Approx 500 charge cycles, under normal conditions of use
Power supply
Battery pack Li-ion 3.7 V 1100mAh
Battery charger
Voltage = 5VDC
Current = 500 mA or higher
Connector = micro USB type B
Dimensions
160x55.2x25mm;
Weight
Central unit 140g (including batteries)
Type of electrical protection
Class II device
Type of electrical protection
BF
Grade of protection against water ingress
IPX1 device, protected against water drops
Safety level in the presence of
inflammable anaesthetic gas, oxygen or
nitrogen
Device not suitable
Conditions of use
Device for continuous use
Storage conditions
Temperature: MIN -20 °C, MAX + 60 °C
Humidity :MIN 10% RH; MAX 95%RH
Transport condition
Temperature: MIN -20 °C, MAX + 60 °C
Humidity :MIN 10% RH; MAX 95%RH
Operating conditions
Temperature: MIN + 10 °C, MAX + 40 °C;
Humidity: MIN 10% RH; MAX 95%RH
Applied norms
Electrical Safety Standard IEC 60601-1
Electro Magnetic Compatibility IEC 60601-1-2
Essential performances (according to EN
60601-1:2007)
Accuracy of spirometry parameters compliant to ATS standard
Measure of the oximetry parameters with accuracy defined in table on page 13
2. FUNCTIONING OF THE SPIROBANK II
2.1 switch on and switch off the device
To switch on the SPIROBANK II push
The first screen shows the manufacturer logo, information of date and hour
set on the device.If no key are touched, after a few seconds the device shows
the main screen.