User's Manual

Spirobank II
Page 18 of 34
Rev.1.1
User manual
Alarms setting
Access to this function allows to setup the parameters linked to the oximetry
First parameter is the alarm intensity: it can be possible to set the type and
volume. Use and to switch from a parameter to the other, then
e are useful to set the desired value: the select icon is the grey one. Press
to change screen. Step following steps allow to set the threshold value
for %SpO2 and BPM. An acoustic alarm will warn the user if the SpO2 and
BPM during a test fall below the minimum threshold or rise above the
maximum threshold of the SpO2 and BPM values previously set.
Use keys and to decrease/increase the values and select moving
the arrow with the keys and .
At the end press to return to the service menu.
WARNING
If the maximum value of a %SpO2/BPM parameter is set lower or equal to the minimum value the setting will not take
effect. The device will emit an acoustic warning and automatically return to the setting of the minimum value.
UNIT format
The voice allows to choose one of the following option:
Imperial (in,lb)
Metric (cm kg)
Select the desired format with or and press ; the selection will be saved automatically and the device will return to the
service menu.
Info firmware
In this menu the user may view information regarding the components version
presents in the device:
Bluetooth version
Bluetooth PIN
Oximeter
After approximately 10 seconds the device will automatically return to the
service menu, otherwise press .
Once all of the items in the service menu have been set it is possible to exit the
menu by pressing .
2.5.1 Turbine calibration
WARNING
The turbine flow sensor does not require calibration, however regular cleaning of the turbine is necessary . The disposable
turbine is check before the application of the external packaging, for this reason it doesn’t requires a periodic calibration. If
a calibration must be performed the following guidelines should be carefully noted.