User's Manual

Spirobank II
Page 28 of 34
Rev.1.1
User manual
The sensor included with the SPIROBANK II is made with latex free material.
4.2 Changing the adhesive wrap sensor
The disposable adhesive tape is made with latex-free material.
Gently remove the used adhesive tape from the sensor and dispose of it.
The back of the sensor has alignment pins. Place the sensor with the alignment pins facing the adhesive part of the tape and align
the pins to the holes on the tape.
Push the sensor so as to insert the pins into the holes of the tape. Lift both the sensor and the tape and check that the pins of the
sensor are properly aligned.
WARNING
It is highly suggested to use a new piece of adhesive tape for each patient, or as required.
4.3 Battery charging
Turn on SPIROBANK II and the following icon will appear on the main screen showing the charge level of the battery pack:
The maximum charge level is displayed with all 6 bars inside the battery.
If only one bar is shown or if the device will not even turn on the battery pack must be recharged in the following manner:
Plug the battery charger into a socket and the battery charger cable into the micro USB
connector of the device; the device in this phase is always turned on
When the charging is complete the battery icon will display all six bars.
At this point disconnect the battery charger from the device.
SpO2
WARNING
It is recommended not to use the device while the battery is charging.
Always disconnect the battery charger from the device when the charge cycle has terminated.
5. PROBLEM SOLVING
PROBLEM
MESSAGE
POSSIBLE CAUSES
REMEDY
SPIROBANK II does
not turn on
\
The battery pack could be discharged
Connect the device to the battery charger.
\
The battery pack has not been properly
inserted in the device
Contact a technical service center
\
The device may have lost its internal software
Connect the device to the PC with the
USB cable and update the internal software;
For more detailed information please
consult the winspiroPRO software user
manual available on line within the
software itself.
Problem when turning
on the device
Error in ram
memory
Recovering
data
Please wait
Memory data within the device has been
damaged
If the data has been restored correctly the
standard turn-on process will complete
itself. If this process does not finalize
contact an authorized technical service
center.
The device turns off and
subsequently turns on
again.
\
An internal error has occurred.
Check on the following website
www.spirometry.com for a more recent
internal software release of the device.
Update the internal software by
downloading the latest release by using the
winspiroPRO software For further
information consult the winspiroPRO
manual available on line within the software
itself.