ANNEX 3 INFORMATION FOR CORRECT USE IN AN ELECTROMAGNETIC ENVIRONMENT
Guidance and manufacturer’s declaration – electromagnetic emissions
The SPIROBANK II is intended for use in the electromagnetic environment specified below.
The customer or the user of the SPIROBANK II should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The SPIROBANK II uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The SPIROBANK II is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The SPIROBANK II is intended for use in the electromagnetic environment specified below. The customer
or the user of the SPIROBANK II should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
In the event of disruption due to ESD during oximetry test, the device
recovers from any disruption within 30 s. (according to ISO 9919).
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for
input/output
lines
Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV
differential mode
±2 kV common
mode
Not Applicable
Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines
IEC 61000-4-
11
<5 % UT
(>95 % dip in
UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in
UT)
for 5 sec
Not Applicable
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.