Instruction manual
Guidance and manufacturer's declaration - electromagnetic emissions 
The Model CP30DW is intended for use in the electromagnetic environment specified below.
The customer or user of the Model CP30DW should assure that it is used in such an environment.
Emissions test
RF emissions
EN 55011
RF emissions
EN 55011
Harmonic emissions
EN 61000-3-2
Voltage fluctuations / 
flicker emissions
EN 61000-3-3
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic environment - guidance
The Model CP30DW uses RF energy only for its 
internal function. Therefore, its RF emissions are very 
low and are not likely to cause interference in nearby 
electronic equipment.
The Model CP30DW is suitable for use in all 
establishments, including domestic establishments and 
those directly connected to the public low-voltage 
power supply network that supplies buildings used for 
domestic purposes.
Guidance and manufacturer's declaration - electromagnetic immunity 
The Model CP30DW is intended for use in the electromagnetic environment specified below.
The customer or user of the Model CP30DW should assure that it is used in such an environment.
Immunity test
Electrostatic 
discharge (ESD)
EN 61000-4-2
Electrical fast 
transient/burst
EN 61000-4-4
Surge
EN 61000-4-5
Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines
EN 61000-4-11
Power frequency 
(50/60 Hz) magnetic 
field
EN 61000-4-8
EN 60601 test 
level
±6 kV contact 
±8 kV air
±2 kV for power 
supply lines
±1 kV for input 
/output lines
±1 kV line(s) to 
line(s)
±2 kV line(s) to 
earth
< 5% 
U
T
(> 95% dip in 
U
T
)
for 0.5 cycle
40% 
U
T
(60% dip in 
U
T
) 
for 5 cycles
70% 
U
T
(30% dip in 
v
T
) 
for 25 cycles
< 5% 
U
T
(> 95% dip in 
U
T
)
for 5 sec.
3 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30%.
Mains power quality should be that of a 
typical commercial or hospital 
environment.
Mains power quality should be that of a 
typical commercial or hospital 
environment.
Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the Model 
CP30DW requires continued operation 
during power mains interruptions, it is 
recommended that the Model CP30DW 
be powered from an uninterruptible power 
supply or a battery.
Power frequency magnetic fields should 
be at levels characteristic of a typical 
location in a typical commercial or 
hospital environment.
Compliance 
level
±6 kV contact 
±8 kV air
±2 kV for power 
supply lines
±1 kV for input 
/output lines
±1 kV line(s) to 
line(s)
±2 kV line(s) to 
earth
< 5% 
U
T
(> 95% dip in 
U
T
) 
for 0.5 cycle
40% 
U
T
(60% dip in 
U
T
) 
for 5 cycles
70% 
U
T
(30% dip in 
U
T
) 
for 25 cycles
< 5% 
U
T
(> 95% dip in 
U
T
)
for 5 sec.
3 A/m
NOTE 
U
T
 is the a.c. mains voltage prior to application of the test level.
Compliance with the following EMC guidance is assured only for the use in Europe (220 - 240 V).