Guidance and manufacturer's declaration - electromagnetic immunity 
The Model CP30DW is intended for use in the electromagnetic environment specified below.
The customer or user of the Model CP30DW should assure that it is used in such an environment.
Immunity test
Conducted RF
EN 61000-4-6
Radiated RF 
EN 61000-4-3
EN 60601 test 
level
3 Vrms
150 kHz to 80 
MHz
3 V/m
80 MHz to 2.5 
GHz
Electromagnetic environment - guidance
Portable and mobile RF communications 
equipment should be used no closer to 
any part of the Model CP30DW, including 
cables, than the recommended 
separation distance calculated from the 
equation applicable to the frequency of 
the transmitter.
Recommended separation distance
d=1.2√P
d=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2.5 GHz
where P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter manufacturer 
and d is the recommended separation 
distance in meters (m).
Field strengths from fixed RF transmitters, 
as determined by an electromagnetic site 
survey,
a
 should be less than the 
compliance level in each frequency 
range.
b
Interference may occur in the vicinity of 
equipment marked with the following 
symbol:
Compliance 
level
3 Vrms
3 V/m
NOTE1. At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2. These guidelines may not apply in all situations. Electromagnetic propagation is affected 
by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which the Model CP30DW is used 
exceeds the applicable RF compliance level above, the Model CP30DW should be observed 
to verify normal operation. If abnormal performance is observed, additional measures may be 
necessary, such as reorienting or relocating the Model CP30DW.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.