user manual

4 - 22
No. 99MBC032A
4.5.2 Outline of each function setup mode
1. Data display unit
If the basic setup mode is entered, the following is displayed.
The function setup number will be flashing in the most significant digit of the upper
display section, and the guidance for the setup item, followed by the setup value, will be
shown to the right of the setup number.
In the lower display section the measurement from the foreground program number will be
displayed.
Function setup number: 1 digit
Setup item guidance: 3 to 8 digits
Setup data: Maximum 10 digits
Unit: 2 digits
Measurement from the
foregroundprogram number
PROG
2. Setting each setup item
Use the numeric keys for setting the setup value, such as an offset value, and use the
and keys for selecting the item, such as the statistical item of the sample
measurement.
Press the
ENT
key to accept and save the setup data. After the setup content has been
accepted, the operation automatically proceeds to the next setup item.
3. Setup values that must meet the large/small relationships
The setup values for GO/NG judgment should meet the following relationships: Abnormal
lower limit < Abnormal upper limit, Lower tolerance limit < Upper tolerance limit, and
Lower limit value < Upper limit value.
If the previously specified setup value needs to be modified to a great extent, it is recom-
mended to first enter the new setup value thats meets the existing large/small relationship
or, for safety, cancel the both sides to 0 then set them again.
4. Confirming the setup contents of each setup item
To confirm the setup contents of each setup item use only the
ENT
key, which does not
affect the setup contents.
5. Terminating the function setup mode
If the
SET
key is pressed while the function setup number is flashing, operation
returns to the ready state.
If the
SET
key is pressed in the setup mode of each setup item, operation returns to
the selection of a function setup number. If the
SET
key is pressed again at this point,
operation returns to the ready state.
If the power is turned off halfway to the setup operation, on-going setup contents will
not be saved in memory. The contents must be set again.