Step 3 Before the touch signal probe is in contact with the measured surface on the
other side, set up the Hold mode using the key.
Step 4 Bring the touch signal probe in contact with the opposite end. The held display
value is the dimension of the workpiece.
Step 5 Quit the touch signal probe measurement function.
IMPORTANT
When measuring with the touch signal probe move the worktable at a traveling speed of
50mm/s or less. If the worktable is moved faster than 50mm/s, the measuring error will
increase.
Leave an interval of at least one second from the time the touch signal probe comes in
contact with the workpiece until the next contact. If the interval is less than one second,
it may be impossible for the counter to detect the measurement point.
Do not displace any axis scale during numeric entry. If displaced during entry for
presetting, etc., the scale unit may be damaged due to an overrun.
Even if the counter exceeds the counting range, Error 30 will not be displayed during
numeric entry. The error message will be displayed after numeric entry has been
completed.
Do not displace any axis scale too much while presetting is being performed.
3.2.9 Halving the display
The KA counter is capable of halving the display value or preset value and displaying the
halved value.
1) Halve the display value for the X-axis.
Halve the currently displayed value, 123.000.
No. 99MBE032A