1.8 Basic adjustments
1.8.1 Adjustment of the rotortype
Before the first operation resp. after each rotor change, you have to settle the respective rotor type.
You find each rotor type in the printed order number on the rotor.
Example:
angle rotor 220.87 V11 rotortype „87“
Switch on the unit and open the lid. Now press simultaneously the keys “lid (7) and “stop” (10). In
the display “rpm/rcf” (A-1) then appears the old settled rotor type, i. e. “87”. With the potentiometer
(1) you can settle the used rotor now. (see figure 2) To adopt the data in the unit please press the
“start” (9) key. Inside the display (A-1) the stroke “store” appears to the confirmation.
Now all important rotor parameters for the centrifuge are stored.
Figure 2
Attention: Set rotor must always match with the used rotor, otherwise you might damage the
unit.
The rotor type can be checked during the run by pressing the key "lid" (7).