2.4 Radius correction
If you use adapters or reducers it could change the centrifugal radius of the respective rotor. In that
case you can correct the radius manually. Please proceed as follows:
Press the key „time“ (6) and the key „prog“ (11) at the same time and hold them.
In the display „time“ (A-3) appears the word „radius“ (M9). By the adjusting knob (1) you can
preselect then the respective radius correction (see table 7, APPENDIX P. VII) in steps of 0,1 cm.
As soon as you have set a radius correction the word „radius“ (M9) appears. This hint is as long
visible as you put the radius correction back to 0 again.
All with number marked passages refer to figure 24