Instruction manual

OPERATION
26 Z216MK_V4.14 © Hermle Labortechnik GmbH
2.5.2 Recall of stored programs
To recall stored programs press the key „prog“ (11) while the lid is already closed. Inside the display
„time“ (A-3) appears „programm --“(M10). With the adjusting knob (1) you pre-select the desired
program number.
In the respective displays there will appear the stored values for that program.
If there is not the right rotor inside the centrifuge for the pre-selected program, in the display
„rpm/rcf“ (A-1) flashes the word „rotor“ (M3). At the same time the word „FALSE“ and the stored
rotor number „22x. xx“ (M4) will flashing by turns.
All with number marked passages refer to figure 28.
Figure 28
2.5.3 Leaving program mode
To leave the program mode just press the key „prog“ (11). Then inside the display „time“ appears
the word „programm".
Set the display to „programm--“ (M10) with the adjusting knob (1).
All with number marked passages refer to figure 28.
2.6 Starting and stopping the centrifuge
2.6.1 Starting the centrifuge
You can start the centrifuge either with the „start“ key (9) or the „quick“ key (8).
By the „start“ key (9) you can start stored runs or runs with manually pre-selected parameters.
When the respective pre-selected running time has ended then the centrifuge will stop automatically.
By the „quick“ key (8) you can start runs, which will last just a few seconds.
By pressing the „quick” key (8) the centrifuge accelerates up to the pre-selected revolution.
In the display „time“ (A-3) the passed running time is indicated from the date of pressing the
„quick” key (8).
By releasing the „quick” key (8) the centrifuge stops and the running time is indicated until the
opening of the lid.
All with number marked passages refer to figure 29