4. TROUBLE SHOOTING
4.1 Error message: Cause / Solution
The error messages are listed to help localize possible errors faster.
The diagnose referred to this chapter may not always be the case, as they are only theoretically
occurring errors and solutions.
Always, please keep us informed about any kind of error occurring, which is not listed in this chapter.
Only through your information we are able to improve and complete this operation manual.
Many thanks in advance for your support.
HERMLE Labortechnik GmbH
4.2 Survey of possible error messages and their solutions
4.2.1 Lid release during power failure (Emergency Lid Release)
In case of power failure or malfunction, the lid of the centrifuge can be opened manually in order to
protect your samples.
Please proceed as follows (see figure 32):
Switch the centrifuge off and unplug the power cord, wait until the rotor stands still (this
may take several minutes)
At the left side of the centrifuge housing there is a plastic stopper. Remove this stopper
and behind it there is a hexagon nut.
Take the delivered box spanner, put him in the hole and lock the box spanner with
hexagon nut (see figure 32).
Now turn the box spanner to the right side (clockwise) up to the limit.
ATTENTION: Just turn to the limit, don´t tighten the nut.
Now open the lid of the centrifuge.
Switch the centrifuge on again, for go on working.
Figure 32