Instruction manual

APPENDIX
VI Z216MK_V4.14 © Hermle Labortechnik GmbH
Table 6: Error messages
Error-No.: Description
1 Imbalance arose
2 Imbalance sensor is defective
4 Imbalance switch has been activated for longer than 5 seconds
8 Transponder in the rotor is defective
11 Temperature sensor is defective
12 Chamber over temperature
14
CLOSE lid
33 Open lid while motor is running
34 Lid contact defective
38 Lid motor is blocked
40 Communication with frequency converter distrubed during start
41 Communication with frequency converter distrubed during stop
42 Short circuit in the frequency converter
43 Undervoltage frequency converter
44 Overvoltage frequency converter
45 Over temperature frequency converter
46 Over temperature motor
47 Over current frequency converter
48 Timeout between control unit and frequency converter
49 Other error frequency converter
55 Overspeed
70 Timeout between controler and RS232 interface
99 Rotor is not allowed in this centrifuge
FALSE Inserted rotor does not exist in the programm
rotor no Rotor is not detected
Leap of speed is too big between two mesaurements