User manual
AM700 Audio Measurement Set User Manual
A-1
Appendix A: Specifications
The performance limits in this specification are valid with the following conditions:
H This instrument must have been calibrated/adjusted at an ambient temperature between
+20° C and +30° C.
H The instrument must be in an environment with temperature, altitude, humidity, and
vibration within the operating limits described in Table A-14 on page A-29.
H The instrument must have had a warm-up period of at least 20 minutes.
H The instrument must have had its signal-path-compensation routine last executed after
at least a 20 minute warm-up period at an ambient temperature within ±5° C of the
current ambient temperature.
Any conditions that are unique to a particular characteristic are expressly stated as part of
that characteristic.
Table A-1: Analog Analyzer Specifications
Capabilities Description
Input Characteristics
Number of Input Channels Two channels: A and B
Input Type Balanced, differential
Input Impedance 100 kΩ per side in parallel with 300 pF (with no
internal loads)
Internal Load Impedance 600 Ω, 150 Ω, or 200 kΩ, software selectable
Vmax Internal Load Approximately 34.65 V peak or 24.5 V
RMS
(sinu-
soidal) at 600 Ω (+30 dBm)
Overload exerted for inputs peaks greater than
+35 V
Maximum Input Voltage 176 V (DC + peak AC)
Note: Installation Category I - Signals must be
isolated from the mains.
DO NOT CONNECT TO MAINS VOLTAGES.