Appendix A: Specifications
The performance limits in this specification are valid with the following conditions:
H This instrument must have been calibrated/adjusted at an ambient temperature between
+20° C and +30° C.
H The instrument must be in an environment with temperature, altitude, humidity, and
vibration within the operating limits described in Table A-14 on page A-29.
H The instrument must have had a warm-up period of at least 20 minutes.
H The instrument must have had its signal-path-compensation routine last executed after
at least a 20 minute warm-up period at an ambient temperature within ±5° C of the
current ambient temperature.
Any conditions that are unique to a particular characteristic are expressly stated as part of
that characteristic.
Table A-1: Analog Analyzer Specifications
Capabilities Description
Input Characteristics
Number of Input Channels Two channels: A and B
Input Type Balanced, differential
Input Impedance 100 k per side in parallel with 300 pF (with no
internal loads)
Internal Load Impedance 600 , 150 Ω, or 200 k, software selectable
Vmax Internal Load Approximately 34.65 V peak or 24.5 V
RMS
(sinu-
soidal) at 600 (+30 dBm)
Overload exerted for inputs peaks greater than
+35 V
Maximum Input Voltage 176 V (DC + peak AC)
Note: Installation Category I - Signals must be
isolated from the mains.
DO NOT CONNECT TO MAINS VOLTAGES.