User manual
Appendix A: Specifications
A-32
AM700 Audio Measurement Set User Manual
Table A-16: Electromagnetic Compatibility (Cont.)
Capabilities Description
EMC Requirements
Immunity, Fast Transients, Common Mode IEC 801-4 1988
No loss of stored data, change to control settings,
or temporary loss of function will occur when the
instrument is subjected to transients as described
below.
Port Peak Tr/Th Rep
(kV) (ns) Frequency
(kHz)
Signal 0.5 5/50 5
&
Control
DC 0.5 5/50 5
Power
AC 1.0 5/50 5
Power
FCC The instrument complies with the radiated and
conducted emissions requirements of FCC 47
CFR, Part 15, Subpart B, for Class A equipment.
To ensure compliance with the above require-
ments, only high-quality shielded interface cable
should be attached to this instrument. High
quality cable is described in the Emissions Speci-
fication.