ProStar Bench Testing Instructions
Table Of Contents
- ProStar Bench Test Instructions
- Tools:
- 1. Getting Started
- 2. Power Up Unit
- 3. Self-Test Routine (Meter Option Only)
- _
- Table 1.Error Summary of Self-Test
- Error
- Description
- Possible Causes
- Solutions
- Notes
- E01
- Battery Switch
- Battery switch in empty 4th position, in between positons, or switch damaged
- Use a screwdriver to ensure that one of the 3 battery types is selected. The switch should click into place
- E03
- Reference or Regulator
- Reference diode or 5V regulator damaged/out of calibaration
- Not a user serviceable failure
- May cause a variety of other errors and failures
- E04
- Array Side
- Damaged Input MOSFETS
- Processor A/D channel damaged
- Array shunt circuit damaged
- Refer to step 4 to confirm failure
- Not a user serviceable failure
- Not a user serviceable failure
- E07
- Load MOSFETs
- Load MOSFETs shorted
- Load Gate drive damaged
- Load +/- jumpered together
- Replace Load MOSFETS
- Not a user serviceable failure
- Check for debris or other objects shorting the load terminals
- No LVD functionality
- No load over-current protection
- E08
- Load Current Zerot
- Load current amplifier damaged
- Load MOSFETs damaged
- Load current circuit damaged
- Not a user serviceable failure
- Replace Load MOSFETs
- Not a user serviceable failure
- E09
- Load MOSFET Open
- Load MOSFETS open
- Load gate drive damaged
- Replace Load MOSFETS
- Not a user serviceable failure
- May reset the processor if short exists on load terminals
- E10
- Internal temp probe open
- Temp sensor damaged
- Temp sensor cut/open
- Reference voltage too low
- Replace temp sensor with a 2N3904 BJT transistor
- Repair / Replace temp sensor with a 2N3904 BJT transistor
- Not a user serviceable failure
- No over-temperature protection
- E11
- Internal temp probe shorted
- Temp sensor damaged
- Temp sensor shorted
- Replace temp sensor with a 2N3904 BJT transistor
- Inspect for debris, find and fix short
- E12
- Remote temp probe out of range
- Temp sensor leads shorted
- Temp sensor leads damaged
- Reference voltage too low
- Remove remote temperature probe, inspect for pinched cable, re-test without probe connected
- Replace probe
- Not a user serviceable failure
- Defaults to on-board temperature sensor. Can not
- If you do not have a probe attached and still get this error, check the connection point on the board where the probe would be soldered for a short. Else, the reference is probably bad.
- E13
- Sense detect
- Long battery leads cause excessive drop
- The “Sense – “ terminal is not connected
- Minimize length of cables from battery to controller, Maximize cable gauge
- Be sure both battery sense lines are securely attached and that they are not damaged
- 4. Check PV / Battery MOSFETs
- 5. Battery Removal Protection
- 6. Putting It All Together
Morningstar Corporation 17 Oct 2001
ProStar Bench Test Instructions
Morningstar Corporation
Last updated: 17 October 2001
Tools:
Phillips screwdriver
Flat-blade screwdriver
DMM(digital multimeter)
Power supply, 20V 2A or similar
Small Ah 12V battery (~10Ah)
Small 12VDC load (eg. Small 1/3A bulb)
NOTE: All equipment ratings and testing values are for 12/24V ProStars. Multiply all voltage ratings and
values by 4 for 48V controllers
1. Getting Started
A few things to do before we begin.
Disconnect unit from system wiring
Open disconnects, switch off breakers
Be careful not to short or damage the controller when removing
Inspect for physical damage to case, heatsink, or protruding components
Remove cover (keep handy to reference terminal positions when wiring during
testing)
Flip the unit over (face down)
Remove the 4 screws that secure the black heat sink to the plastic enclosure
Carefully remove the plastic case
Not necessary to remove external temperature probe if installed
2. Power Up Unit
The controller’s power up routine indicates that the processor is operating and that the controller has not
had a catastrophic failure.
Procedure
Attach a small power supply as battery:
Page 1 of 11 Rev.1.0