Important Information Regarding Electromagnetic Compatibility(emc) 43
Guidance and Manufacturer’s Declaration
– Electromagnetic Immunity
The thermometer is intended for use in the electromagnetic environment
specified below. The customer or user of this thermometer should ensure that
it is used in such an environment.
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
– Guidance
Conducted
RF IEC
61000-4-6
Radiated RF
IEC 61000-4-3
Not
applicable
3 V/m
80 MHz to
2.5 GHz
3 V/m When used near any part of
the thermometer (including
cables), portable and mobile
RF communications equipment
should not be used any closer
than the recommended
separation distance calculated
based on the equation
appropriate to the frequency of
the transmitter.
Recommended separation
distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where “P” is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and “d” is the
recommended separation
distance in meters (m). Field
strengths from fixed RF
transmitters as determined by
an electromagnetic site survey