Datasheet

Uz210_02c_oi_e.doc / Dez-15 Page 20 / 27
5.7. Linearization Setting
No.
Linearization Table
Range
Default
Ser.
050
First interpolation point (x0, original value)
E6
051
First interpolation point (y0 as substitution for x0)
052
Second interpolation point (x1, original value)
-10 000 … +10 000
0
053
Second interpolation point (y1 as substitution for x1)
etc. ---->
080
Last interpolation point (x15, original value)
081
Last interpolation point (y15 as substitution for x15)
H7
5.8. Hints for Use of the Linearization Function
The drawings below explain the difference between the settings
„Linear Mode“ = 1 and „Linear Mode“ = 2:
x
y
(x0)= 0
(y0)= 0
Linearization Mode
Setting "1"
x
y
Linearization Mode
Setting "2"
(x0)= -1000
(y0)= 900
(x8)= 0
(y8)= 750
(x15)= +1000
(y15)= - 600
*)
(x15)= 1000
(y15)= 800
The x-values are to determine which originally measured input value should be
substituted by another value
The corresponding y value defines a new value for replacement of the previous
x value (e.g. the value x3 will be modified into y3)
Values between two interpolation points will be reproduced by straight lines (linear
interpolation)
x-registers must use continuously increasing settings, i.e. P0(x) must have the lowest
and P15(x) must have the highest setting
Independent of all other settings the acceptable range for x values and
y values is always from -10 000 to +10 000.
For measuring values outside of the defined linearization range, please note:
- if a measuring value is lower than x0, the linearization result will always be y0.
- if a measuring value is higher than x15, the linearization result will always be y15.