User's Guide

POLI User’s Guide
24
6. Calibration and Test
6.1 Manual Alarm Test
Under normal operation mode and non-alarm conditions, the buzzer (audible alarm),
vibration, visible alarms, and backlight can all be tested anytime by pressing [+/OK]. If
any alarm does not respond, check the alarm settings to make sure all alarms are enabled
(LED, Buzzer, Vibrator under Config mode/Alarm/Alarm Device should be all
Enabled’). If any alarms are enabled but not functional, the instrument should not be
used. Contact mPower’s technical support.
6.2 Bump Testing and Calibration
mPower recommends that a bump test be conducted periodically on the POLI. The
purpose of a bump test is to ensure that the instrument’s sensors respond to gas and all
the alarms are enabled and functional.
The POLI multi-gas detector must be calibrated if it does not pass a bump test, or
at least once every 180 days, depending on use and sensor exposure to poisons and
contaminants.
Calibration intervals and bump test procedures may vary due to national and local
legislations or regulations.
A bump test can be performed manually. When a bump test is done manually, the
instrument makes a pass/fail decision based on sensor performance, but the user still has
the responsibility to make sure all the alarms are enabled and functional.
6.2.1 Bump (Functional) Testing
The same gas is used for a bump test as for calibration. A constant-flow regulator
producing 0.5 to 1.0 liters per minute should be used.
Pumped models require a T calibration tube, as illustrated below.