Instruction manual

concentration of vapors. In general, combustible gases with
flash points above 100
o
F do not give off enough vapors, at
ambient temperatures, to be detected.
When using a sampling pump and sample lines, the user must
wait the appropriate time for gas to be drawn through the
sample system to the sensors. Typical transport times are a
maximum of 0.3 seconds per foot of sample line.
When sampling over liquids, the end of the sampling line must
not touch the surface of the liquid. Otherwise, liquids may enter
the instrument, block sample gas from entering the line, and
cause a false reading to occur. In addition, internal damage to
the instrument may result.
Obstruction of the sensor holes in the instrument case causes
erroneous readings. These holes must be kept open at all
times. Do not use compressed air to clean the sensor holes;
excessive pressure at the face of the sensors could damage
them.
Do not use MSA Lead Inhibitor Filters with this instrument. Loss
of sensitivity can result.
Battery packs must be recharged in a non-hazardous location
free of combustible gases and vapors.
A calibration check must be performed before each day’s use to
verify that the instrument is operating properly and readings are
accurate (or more frequently if the instrument is subjected to
significant physical shock or high levels of contaminants). If the
readings are not within the specified limits, the instrument must
be recalibrated before use. If no calibration check is performed,
inaccuracies in gas readings may not be detected. See the
FiveStar Alarm Technical Manual, Chapter 2, for calibration
check procedure.
Use only genuine MSA replacement parts when performing any
maintenance procedures described in this manual. Substitution
of components can seriously impair instrument performance,
alter intrinsic safety characteristics, or void agency approvals.
Repair or alteration of the FiveStar Alarm beyond procedures in
this manual or by anyone other than a person authorized by
MSA could cause the instrument to fail to perform properly.
FAILURE TO FOLLOW THE ABOVE WARNINGS CAN RESULT
IN SERIOUS PERSONAL INJURY OR DEATH.
Instruction Manual Chapter 1, Safety and General Limitations
FiveStar Alarm 1-3