Instruction manual

Persons responsible for the use of the
FiveStar Alarm must determine
whether or not the Fresh Air Setup
option should be used. The user’s
abilities, training and normal work
practices must be considered when
making this decision.
• When the (FIGURE 2-19) display
appears, the FiveStar Alarm is
ready for its Fresh Air Set Up.
To Proceed With Fresh Air
Setup:
1. Press the RESET (YES) button;
the display reads:
ADJUSTING ZEROES
PLEASE WAIT
a. When the (FIGURE 2-20)
display appears:
• Alarm sounds
• Lights flash
1) Push the RESET (OK) button.
2) Make certain the FiveStar Alarm is in fresh air;
move to another location, if necessary. Allow the
FiveStar Alarm to warm up for a few minutes to allow
the sensors to stabilize.
Turn FiveStar Alarm OFF and then back ON again.
3) If the FiveStar Alarm cancels the Fresh
air Setup request again, calibration adjustments may be
required. Report to the person responsible for
FiveStar Alarm maintenance.
Do not use the instrument for protection.
b. When the Fresh Air Setup is completed:
• Instrument enters the Exposure display page,
displays gas readings, and is ready for use.
To Bypass The Fresh Air Set Up:
Press the PAGE (NO) button, or wait five seconds.
Figure 2-19.
Fresh Air Set-up
Figure 2-20.
FAS Canceled/OK
Chapter 2, Using the FiveStar Alarm Instruction Manual
2-8 FiveStar Alarm