Instruction manual

• The instrument will continue to operate until the power
is turned OFF or the battery condition is at BATTERY
SHUTDOWN level.
Approximate Voltage Readings for
Replaceable Battery Pack Alarms
WARNING 3.3 VOLTS
SHUT-DOWNS 3.1 VOLTS
• BATTERY SHUTDOWN: the battery is no longer able to
operate the instrument, and:
• BATTERY SHUTDOWN appears in place of the
Exposure Display Page. Horn sounds continuously and
cannot be reset.
• Alarm lights flash and the horn sounds intermittently.
• No other pages can be viewed.
• After approximately five minutes, the instrument shuts
down automatically.
"
WARNING
When the Battery Shutdown condition sounds, stop using
the instrument. It cannot alert you of potential hazards
because it does not have enough power to operate properly.
You must:
1. Leave the area immediately.
2. Turn OFF the instrument if it is ON.
3. Report to the person responsible for maintenance.
Replace or recharge the battery pack.
If you do not follow this procedure, you could be injured
or killed.
"
WARNING
For Replaceable Battery packs, replace batteries when the
"Battery Low" or "Battery Shutdown" alarms occur. When
replacing lithium or alkaline batteries, replace ALL batteries
with fresh ones at the same time. Do not mix battery types
or new and partially-discharged batteries. If the batteries
are improperly replaced or improperly mixed, the "Battery
Low" and "Battery Shutdown" alarms may fail to function,
which could result in serious personal injury or death.
Instruction Manual Chapter 2, Using the FiveStar Alarm
FiveStar Alarm 2-11