Instruction manual

1995" would read as shown (FIGURE 2-25) at 7:06 p.m.
Measuring Gas Concentrations
Combustible Gases (COMB)
The FiveStar Alarm detects combustible gases in the atmosphere.
The Alarms sound when concentrations reach:
Alarm setpoint, or
100% LEL (Lower Explosive Limit), or
5% CH
4
(Methane by volume)
When the combustible gas indication
reaches the Alarm Setpoint:
Alarm sounds
Alarm lights flash
Press the RESET button to
silence the alarm. (The alarm
will stay silent if the alarm
condition has cleared.)
Concentration of gas flashes in the
display (FIGURE 2-26).
When the combustible gas indication reaches 100% LEL or 5%
CH4 of the combustible gas:
Alarm sounds
Alarm lights flash
This alarm cannot be reset with the RESET button.
The LockAlarm circuit locks the combustible gas reading and
alarm if the gas reading exceeds 100% LEL or 5% methane.
OVER appears on the display.
"
WARNING
If the OVER alarm condition is reached, you may be in a
life-threatening situation; there may be enough gas in the
atmosphere for an explosion to occur. In addition, any rapid
up-scale reading followed by a declining or erratic reading can
also be an indication that there is enough gas for an explosion.
If either of these indications occur, leave and move away from
Figure 2-26.
Combustible Gas Alarm
Flag
Instruction Manual Chapter 2, Using the FiveStar Alarm
FiveStar Alarm 2-13