Instruction manual

Setting a Tag
1. Press the PAGE button until the (FIGURE
2-35) screen appears:
2. Press the RESET (YES) button.
• A screen showing the last saved data
tag appears (FIGURE 2-36).
3. Press OK to accept the alphanumeric tag
or press EDIT to change the tag.
• If no button is pressed within 10
seconds, this data tag is
automatically stored.
• The display returns to the
Measure Gases Page.
a. If you press EDIT, a
screen similar to (FIGURE
2-37) appears.
b. Press PAGE or RESET
(up or down arrows) to edit
the first character.
c. Press ON/OFF (NEXT) to
proceed to the next character
in the tag.
d. Repeat Steps (b) and (c).
• Tag is automatically
recorded when last
character is entered.
• The (FIGURE 2-38)
screen appears:
Autocalibration
The Autocalibration sequence resets the
instrument zeroes and adjusts the sensor
calibration for known concentration of
calibration gases. Autocalibration is standard
on FiveStar Alarms when factory-shipped with
any combination of the following sensors:
Figure 2-35.
Setting Tag Page
Figure 2-36.
Last Data Tag Page
Figure 2-37.
Editing Tag Page
Figure 2-38.
Recording Tag Page
Chapter 2, Using the FiveStar Alarm Instruction Manual
2-20 FiveStar Alarm