Instruction manual

Setting a Tag
1. Press the PAGE button until the (FIGURE
2-35) screen appears:
2. Press the RESET (YES) button.
A screen showing the last saved data
tag appears (FIGURE 2-36).
3. Press OK to accept the alphanumeric tag
or press EDIT to change the tag.
If no button is pressed within 10
seconds, this data tag is
automatically stored.
The display returns to the
Measure Gases Page.
a. If you press EDIT, a
screen similar to (FIGURE
2-37) appears.
b. Press PAGE or RESET
(up or down arrows) to edit
the first character.
c. Press ON/OFF (NEXT) to
proceed to the next character
in the tag.
d. Repeat Steps (b) and (c).
Tag is automatically
recorded when last
character is entered.
The (FIGURE 2-38)
screen appears:
Autocalibration
The Autocalibration sequence resets the
instrument zeroes and adjusts the sensor
calibration for known concentration of
calibration gases. Autocalibration is standard
on FiveStar Alarms when factory-shipped with
any combination of the following sensors:
Figure 2-35.
Setting Tag Page
Figure 2-36.
Last Data Tag Page
Figure 2-37.
Editing Tag Page
Figure 2-38.
Recording Tag Page
Chapter 2, Using the FiveStar Alarm Instruction Manual
2-20 FiveStar Alarm