Instruction manual

Table 2-1. Autocalibration Allowable Sensors & Required Calibration
Cylinders
AUTOCALIBRATION
ALLOWABLE
SENSORS
EXPECTED GAS
CONCENTRATION
THREE-GAS
CYLINDER
(MSA P/N 10010162)
FOUR-GAS
CYLINDER
(MSA P/N 804770
OR P/N 711058)
COMBUSTIBLE
58% ✔✔
OXYGEN
15% ✔✔
CARBON
MONOXIDE
300 PPM ✔✔
HYDROGEN
SULFIDE
10 PPM
If any other sensors are installed in the
instrument, the feature is automatically disabled.
The Autocalibration feature can also be disabled
by using the MSA FiveStar LINK
TM
.
To access Autocalibration, press and hold
the RESET button for three seconds.
(FIGURE 2-39) display appears.
Press the RESET (YES) button to start the
Autocalibration sequence.
The instrument display indicates
that it is adjusting zeroes.
NOTE: During the Autocalibration zero
procedure, the limits on the
zero adjustments present
during the Fresh Air Setup
option are eliminated.
NOTE: For autocalibration, all
gases must be in one cylinder.
See TABLE 2-1 for allowable
cylinders and concentrations.
After zeroes are adjusted, the
instrument prompts the user to
apply calibration gas (FIGURE
2-40) and to cycle through the gas
readings (FIGURE 2-41)
one-at-a-time for 90 seconds.
If the Autocalibration sequence
has passed, the instrument briefly
displays a screen reminding the
user to remove the calibration gas
Figure 2-39.
Auto Cal? YES/NO
Figure 2-40.
Apply Cal Gas
Figure 2-41.
Example of a
Combustible Gas
Reading
Instruction Manual Chapter 2, Using the FiveStar Alarm