Instruction manual

Nitric Oxide Sensor Specifications
Table 4-9. NITRIC OXIDE (appropriate models only) -
Typical Performance Specifications
RANGE
100 ppm
RESOLUTION
1 ppm
REPRODUCIBILITY
+2 ppm or +10% of reading, whichever is greater
RESPONSE TIME
25 seconds*
*See TABLE 4-1, NOTE 1.
Table 4-10. NITRIC OXIDE- Cross Reference Factors for FiveStar
Calibration Using Calibration Cylinder (P/N 812144) Set to 50 ppm NO
NOTE: Data is presented as the indicated output in ppm,
which would result from the application of 100 ppm of the test gas.
TEST GAS EQUIVALENT PPM
Carbon Monoxide (CO)
0
Hydrogen Sulfide (H
2
S)
1
Nitrogen Dioxide (NO
2
)
30 +5
Sulfur Dioxide (SO
2
)
0
Ammonia (NH
3
)
0
Carbon Dioxide (CO
2)
0
Chlorine Dioxide (ClO
2
)
75
Instruction Manual Chapter 4, Performance Specifications
FiveStar Alarm 4-7