Specifications

GB
OPERATION MSA
16 ALTAIR PRO - Operating Manual
High Alarm Setpoint for three seconds, "HI" and "ALARM" icons turn ON.
(8) To change the high alarm (HI), press the button when "HI" and "ALARM"
displays.
"
HI", "ALARM", "SET" and "?" are displayed.
(9) By pressing the button, increase the value for the high alarm, until the
desired value is reached.
On obtaining the maximum value, the display jumps back and begins to
count upwards again from the minimum value.
(10) Release the button and wait 3 seconds.
Short Term Exposure Limits "STL" displays; then, the STEL setpoint displays
for three seconds.
(11) To change the STEL setpoint, press the button when the setpoint displays.
"
ALARM", "SET" and "?" are displayed.
(12) By pressing the button, increase the value for the STEL alarm, until the
desired value is reached.
On obtaining the maximum value, the display jumps back and begins to
count upwards again from the minimum value.
(13) Release the button and wait 3 seconds.
Time Weighted Average, "TWA" displays; then, the TWA setpoint displays for
three seconds.
(14) To change the TWA setpoint, press the button when the setpoint displays.
"
ALARM", "SET" and "?" are displayed.
(15) By pressing the button, increase the value for the TWA alarm, until the
desired value is reached.
On obtaining the maximum value, the display jumps back and begins to
count upwards again from the minimum value.
(16) Release the button and wait 3 seconds.