Specifications
GB
OPERATION MSA
24 ALTAIR PRO - Operating Manual
3.5. Displaying the Instrument Data - excluding the Steel Mill Version
A single, quick press on the button turn ON the backlight for 10 seconds. The
Information page can be accessed by pushing the button for about one second.
The instrument beeps twice.
There are following modes:
1) Test gas mode
2) Minimum Oxygen concentration ("
LO") - for oxygen only
3) Peak Reading
Toxic ("HI")
Oxygen ("HI")
The Peak/Hi and Min/Low values can be cleared.
When this page is displayed, press the button to clear.
"
CLR" is displayed
Value is deleted
4) Short Term Exposure Limit ("
STL")
The STL reading displays that the instrument calculated since power-ON.
When this page is displayed, press the button to clear.
"
CLR" is displayed
Value is deleted
STL value is automatically reset to zero when the instrument is turned ON.
STEL value is calculated over a 15-minute exposure.
The following formula is used to calculate the STEL value:
((Minute #1 PPM value) + (Minute #2 PPM value) +...
+ (Minute #15 PPM value)) / (15 minutes) = PPM STEL value
If the instrument is ON for less than 15 minutes:
the balance of the minute PPM values is set to zero and
the total is divided by 15 minutes
5) Time Weighted Average ("TWA")
The TWA reading displays what the instrument has calculated since power-
ON. When this page is displayed, press the button to clear.
"
CLR" is displayed
Value is deleted
TWA value is automatically reset to zero when the instrument is turned ON.
TWA value is calculated over an eight-our exposure.
The following formula is used to calculate the TWA value:
Sum of 1-minute gas readings/480 minutes = PPM TWA value
If the instrument is ON for less than 8 ours (480 minutes), the balance of the
minute ppm values is set to zero.