Specifications
GB
MSA OPERATION
ALTAIR PRO - Operating Manual 33
(7) If calibration fails, check:
whether the correct test gas was used,
whether the test gas cylinder is empty or the date has expired,
whether the test gas hose was connected to the sensor,
whether the flow governor is set to 0.25 l/min.
(8) If necessary, repeat steps (1) to (7).
"OK" is displayed. Otherwise ("
ERR"), the instrument must be taken out
of use.
(9) Conduct a self-test to confirm the calibration.
If the self-test is successful, a "" is displayed.
Factory Set Default Calibration / Bump Test Values
INSTRUMENT TYPE BUMP TEST GAS CALIBRATION GAS
CO 60 ppm 60 ppm
H
2
S 40 ppm 40 ppm
O
2
and O
2
-R <19% * 20.8%
SO
2
10 ppm 10 ppm
NO
2
10 ppm 10 ppm
NH
3
25 ppm 25 ppm
PH
3
0.5 ppm 0.5 ppm
HCN 10 ppm 10 ppm
Cl
2
10 ppm 10 ppm
ClO
2
2 ppm Cl
2
0.8 ppm ClO
2
**
* The O
2
bump test can also be performed by exhaling on the sensor inlet for approximately three to
five seconds.
** MSA recommends calibration of the ClO
2
ALTAIR PRO Detector using a ClO
2
generator for most
accurate calibration. This instrument has a cross-sensitivity factor to Cl
2
of:
2 ppm Cl
2
equals approximately 0.8 ppm ClO
2
.