User`s manual
120 cHAPTER 6: Computer Interface Operation
Model 335 Temperature Controller
MODE Remote Interface Mode Command
Input MODE <mode>[term]
Format n
<mode> 0 = local, 1 = remote, 2 = remote with local lockout.
Example MODE 2[term] places the Model 335 into remote mode with local lockout.
MODE? Remote Interface Mode Query
Input MODE?[term]
Returned <mode>[term]
Format n (refer to command for description)
MOUT Manual Output Command
Input MOUT <output>,<value>[term]
Format n, +nnnnn[term]
<output> Specifies output to configure: 1 or 2.
<value> Specifies value for manual output.
Example MOUT 1,22.45[term] Output 1 manual output is 22.45%.
Remarks Manual output only applies to outputs in Closed Loop PID, Zone, or Open Loop modes.
MOUT? Manual Output Query
Input MOUT? <output>[term]
Format n
<output> Specifies which output to query: 1 or 2.
Returned <value>
Format + nnnnn[term] (refer to command for description)
OPST? Operational Status Query
Input OPST? [term]
Returned <bit weighting> [term]
Format nnn
Remarks The integer returned represents the sum of the bit weighting of the operational sta-
tus bits. Refer to section 6.2.5.2 for a list of operational status bits.
OPSTE Operational Status Enable Command
Input OPSTE <bit weighting> [term]
Format nnn
Remarks Each bit has a bit weighting and represents the enable/disable mask of the corre-
sponding operational status bit in the Operational Status Register. This determines
which status bits can set the corresponding summary bit in the Status Byte Register.
To enable a status bit, send the command OPSTE with the sum of the bit weighting for
each desired bit. Refer to
section 6.2.5.2 for a list of operational status bits.
OPSTE? Operational Status Enable Query
Input OPSTE?[term]
Returned <bit weighting> [term]
Format nnn (Refer to section 6.2.5.2 for a list of operational status bits)