Instruction Manual

a) Do this by changing the drop-down selection box to Tuning in the Station Manager application
main window.
b) Enter the password Tuning (case sensitive) when prompted.
2. Set Displacement tuning parameters.
a) Open the Tuning: Ch1 Displacement window by selecting Station Manager > Display > Station
Setup > Channels > Ch1 (or other depending on your configuration) > Displacement, clicking
the Tuning Fork icon, and selecting the Adjustments tab.
b) Use the following parameters for this hard specimen test:
P Gain — 40
I Gain — 1
D Gain — 0
F Gain — 0
F2 Gain — 0
Proportional FF Gain — 0
FL Filter — Set to maximum positive value
Note: The parameters shown are for the Example HCF Test using a hard specimen. For
other tests, we recommend initial settings based on known good settings from tests similar
to the one you are setting up to run. If you are unsure of what these initial settings should be,
contact MTS technical support.
3. Set Force tuning parameters.
70 | MTS Landmark Test System Operation
Running the Example HCF Test