Owner's manual

Series 322 Load Unit
86
Maintenance
244 Actuator: Maintenance
The series 244 Actuator is designed for extended periods of operation without
extensive maintenance requirements.
A summary of the routine maintenance procedures is listed below:
Weekly Clean exposed areas of the actuator piston rod with a clean, dry, lint free rag. If
the actuator is continually exposed to a dirty operating environment, clean the
piston rod on a daily basis.
Monthly Inspect actuator piston rod and seals for excessive wear and/or leakage. Small
scratches in the axial direction of the piston rod or polishing of the rod surface is
considered normal operating wear.
Yearly Change actuator seals if necessary. Actuator assemblies can require more or less
frequent seal changes depending on usage. External oil leakage and/or decreased
performance are indicators of seal wear. Seal replacement is considered a service
procedure; contact you MTS Service Engineer for additional information.
298 HSM: Maintenance
The series 298 Hydraulic Service Manifold requires the filter be changed
periodically.
Component Identification
Pressure
Accumulator
Return
Accumulator
Inlet
Filter
Dirty
Filter
Indicator