Software Owner's manual

DescriptionItem
calibrate the transducer/AC conditioner pair in the new clock
mode and run your tests in the same clock mode.
System Rate: 512 (single-box Aero systems), 1024, 2048, 4096 and
6144 Hz (binary).
Note:
The System Rate must be an integer multiple of the Medium
System Rate.
Medium System Rate: Must be an integer multiple of the Low System
Rate.
Note:
The Low System Rate is always set at 25.6. The Medium
System Rate is typically 256.
Mezzanine: describes the mezzanine cards that are installed on the
I/O Carrier board.
49x.40 Mezzanines tab
Model 494.41 Single-Station System Board Settings
The Model 494.41 board is used with FlexTest 40 controllers.
DescriptionItem
The System Board defines digital and analog inputs and outputs,
and hydraulic power unit (HPU) and hydraulic service manifold
(HSM) functionality.
494.41 System Board circuit
board icon
Slot: 4.Properties tab
Slot 4 in the FT-40 chassis is reserved for the system board.
System Rate: 1024, 2048, 4096 and 6144 Hz (binary).
The System Rate must be an integer multiple of the Medium
System Rate.
Medium System Rate: Must be an integer multiple of the Low
System Rate.
The Low System Rate is always set at 25.6. The Medium System
Rate is typically 256.
Low System Rate: 25.6.
Name: Describes this resource as it will appear in Station Builder
resource lists when opened against an .hwi file. The name
includes the connector number (for example: “Dig In 1-J54”).
Digital Input/Digital Output icons
MTS Series 793 Utility Software 33
HWI Hardware Settings