Owners manual

14
CARE AND MAINTENANCE INSTRUCTIONS
Federal Regulations require Coast Guard approved PFDs that are
legibly marked with the Coast Guard approval number, are in good
and serviceable condition, and are the correct size for each person on
board. It is the responsibility of the operator to be sure this PFD is
in serviceable condition. To be considered serviceable, this PFD shall
not exhibit deterioration that could diminish its performance, such as
broken or deformed hardware, detached webbing, rotted structural
components, air leaks or nonfunctional inflators. Unless worn inflated,
this PFD must also be properly armed with a cylinder installed in
the inflator, a yellow bobbin (water sensing element) installed for
MD2016, and service indicator showing green. See Component
Inspection Instructions. Failure to follow these instructions rigorously
could result in a non-functioning PFD causing injury or death.
Check the following at the intervals listed below.
1) Before Each Use:
i) Examine the service indicator through the window panel,
ensuring it is green. If the indicator is red, the mechanism
requires service. See Re-arming Your PFD (MD2014) on page
8 or see Re-arming Your PFD (MD2016) on page 10.
ii) Ensure the manual pull-tab is accessible.
iii) Visually examine your Inflatable PFD for damage or excessive
abrasion, wear, tear or contamination. Particular attention
must be paid to the stitching, straps and hardware. If in
doubt, send it to your Mustang dealer for evaluation and/or
servicing.
2) Every Two (2) Months:
i) Leak Test: The Inflatable PFD should be tested for general
leakage by orally inflating your Inflatable PFD until firm and
let stand for 16 hours. A leaking Inflatable PFD will not hold
its firmness and should be replaced. If your Inflatable PFD
leaks, take it to your Mustang dealer for evaluation and/
or servicing. Repack the Inflatable PFD as outlined in the
Repacking section on page 13.
ii) Oral Inflation Valve Test: To check the oral inflation valve,
fully inflate the Inflatable PFD using the oral inflator and
hold valve under water. If bubbles appear, deflate and
re-inflate to test again. Should the leak persist, take your