Users Manual
SENSE MCE IBC Operating Instruction 
Rev. D, 2020-12-15  Release    Page 31 of 46 
© MYNXG Product GmbH, 2020. All rights reserved 
5  Electrical Standards and Specifications 
IMPORTAN T NOTICE  
Absolute  Maximum  Ratings  are  those  values  within  damages  on  the 
device will not occur. Any operation outside these maximum ratings may 
cause damages on the device. Recommended Operating Conditions are 
conditions  under  which  operations  of  the  device  are  intended  to  be 
functional. Typically, values correspond to temperature TJ = 25 
°
C. Unless 
otherwise stated, the following conditions apply for supply voltage: Vin = 
5 VDC through Wireless Charger. 
5.1  Recommended Operating Conditions 
The  recommended  operating  conditions  specify  the  values  under  which  the  device 
operationally meets the specifications (Table 1). The battery can provide continuously 3A or 
10A for less than 2ms in temperatures below 45 
°
C. 
CAUTION 
Battery discharge in ambient temperatures lower than 0°C will result in 
limited performance in current output and usable capacity. 
Table 1: Recommended operating conditions. 
Symbol 
Parameter 
Min. 
Typ. 
Max. 
Unit 
Vbat 
Battery Voltage on the battery connector 
3.0 
3.7 
4.2 
V 
Ibat 
Battery maximum continuous discharge current 
0.44 
3
5
A 
Ibatpk 
Battery maximum peak discharge current 
10
6
Vchg 
Wireless Charger Voltage 
4.5 
4.9 
5.0 
V 
Imax 
Wireless Charger maximum current 
1.20 
A 
T
discharge
The  continuous  discharging  temperature 
(limited by the connector in maximum value) 
-20 
- 
55 
°
C 
T
charge
The continuous charging temperature (limited by 
the connector in maximum value) 
0 
- 
35 
°
C
5
 -20°C < T < 55°C (limited by connector).  
6
 0°C < T < 35°C for time period < 2ms 