Appendix 3 Technical Specification
DEVICE TYPE AND MEASUREMENT METHOD
Device type
Myometer for in vivo non-invasive measurement of soft
biological tissues
Measuring
method
Based on exerting a damped mechanical oscillation of soft
biological tissue by
using an exterior, light, quick-release mechanical impulse under
constant pre-load.
Measurable
parameters
State of tension:
- Natural oscillation frequency [Hz] - characterizing Tone or Tension
Biomechanical properties:
- Logarithmic Decrement of natural oscillation characterizing
Elasticity
- Dynamic Stiffness [N/m]
Viscoelastic properties:
- Creep (Deborah number)
- Mechanical Stress Relaxation time [ms]
Precision of
parameters:
Coefficient of
Variation
Natural oscillation frequency - 0,6%
Logarithmic Decrement - 1.8%
Dynamic Stiffness - 1.7%
Mechanical Stress Relaxation time - 1.2%
Creep - 1.5%
Raw signal
processing
Based on proprietary algorithm
MEASUREMENT SYSTEM
Measuring
mechanism
System of frictionless translation motion, driven by solenoid
actuator
Measurement
sensor
3-axis Digital Acceleration Sensor,
Measurement range ±8G, Resolution 11 bits,
Sensitivity mg 13-bit;
Sensitivity range due to temp. ±0.01%
Measurement
probe
3 mm (diameter) polycarbonate probe
Replaceable by user
Probe positioning
Shock protected servo mechanism