Operation Manual
MyotonPRO User Manual
Page 17 of 108
2. It is recommended
6
no to
exceed measurement angles
over 100⁰ from vertical position
as shown on the figure.
In the picture:
A - The Device side view
B - The cross-section of the
Object being measured
GV - Gravity vector
3. In case of conducting
measurements with the Device
positioned horizontally, the
Device must be in flat position,
the screen is either facing
directly up or down. Tolerance
of flat position is +/- 10⁰.
In case of the Device is inclined
more than 10⁰, the warning
notice Rotate! will be displayed.
In this case the Device must be
turned back to horizontal
position with screen facing
directly up or down.
6
It is not recommended to conduct measurements with the Probe facing up (i.e. opposite to the gravity vector), because
in this position the natural oscillation of the Object is significantly influenced by the gravity force.