User guide
CELL
Self-conditioning of the built-in electrodes occurs each time 
the button is pressed with a sample in the cell cup. This ensures 
consistent results each time. With some samples a small 
downward swing of the pointer is a result of this conditioning 
action. This action is powerful and removes normal fi lms of oil 
and dirt. However, if very dirty samples—particularly scaling 
types—are allowed to dry in the cell cup, a fi lm will build up. This 
fi lm reduces accuracy. When there are visible fi lms of oil, dirt 
or scale in the cell cup or on the electrode, scrub them lightly 
with a small brush and household cleanser such as Comet™ or 
Windex™. Rinse out the cleanser and the meter is now ready 
for accurate measurements.
WATER INSIDE THE METER
Your Myron L PoolMeter is a rugged instrument and will 
withstand water exposure around its cell, meter movement, and 
switches. However, care should be taken to keep water from 
leaking in around the bottom cover. It is not sealed in order to 
prevent condensation from forming.
If water is relatively clean (i.e., tap water or better), and there 
are only a few drops inside the meter, dry it as described below. 
Large amounts of water, or corrosive or very dirty solutions will 
almost certainly damage the meter movement or electronics. 
Such meters should be returned to the Myron L Company for 
repair.
TO DRY YOUR METER:
1. Shake excess water out of the inside of the meter.
2. Dab the exposed surfaces dry with an absorbent cloth or 
tissue. Avoid getting any water into the Calibration Control or 
either of the switches.
3. Air dry the meter in a warm area with the bottom cover off. 
Allow several hours for thorough drying.
 If the water entered through a leak in the case or cell, or if the 
instrument shows erratic readings or other unusual behavior, 
return it to the Myron L Company for service.
SERVICE
Any service required other than battery replacement, cleaning 
or calibration must be referred to the Myron L Company. 
Additionally, if you have any questions about this instrument, 
its use, or a particular application, please contact the Myron L 
Company directly (see back cover for contact information).
ACCESSORIES
STANDARD SOLUTIONS
442™: Unless otherwise specifi ed, your PoolMeter has been 
factory calibrated with the TDS 442 Standard Solution. The 
442 Standard Solutions consist of the following salt ratios: 40% 
sodium sulfate, 40% sodium bicarbonate, and 20% sodium 
chloride. This salt ratio has conductivity characteristics closely 
matching natural waters and was developed by the Myron L 
Company almost four decades ago.
Sodium Chloride: For every ppm 442 Standard Solution, 
there is a ppm sodium chloride (NaCI) solution that will have 
the same conductivity. The parts per million of the equivalent 
NaCI solution is on each Standard Solution label. Instruments 
calibrated to NaCI standards are set using equivalent NaCI 
values.
All Myron L Company Standard Solutions are traceable to the 
National Institute of Standards and Technology (NIST) and are 
within 1.00% of reference solutions. The concentrations of the 
reference solutions are calculated from data in the International 
Critical Tables, Vol. 6.
All Myron L conductivity Standard Solution bottle labels 
show three values: ppm 442, ppm sodium chloride, and 
conductivity in micromhos.
RECOMMENDED STANDARD SOLUTIONS
SOLUTION TYPE  ppm value 442™  ppm value NaCI
512T5D Calibration
442-3000 3000 2027
RE-10 Calibration
442-30,000  30,000/30 ppt  18,235/18.24 ppt
RANGE EXTENDER
Model RE-10
The RE-10 Range Extender is a useful accessory for testing 
high conductivity/ppm solutions beyond the normal range of 
your PoolMeter. Inserting the RE-10 into a sample-fi lled cell 
cup increases the maximum range ten times. Use the Range 
Extender whenever the reading is off the scale.
USING THE RE-10
1. To rinse, fi ll the cell cup three times while inserting the Range 
Extender.
2. Fill the cell cup with your sample. Push the Range Extender 
into the cell cup, seating the O-ring seal.
3. Use and read the PoolMeter in the normal manner. Multiply 
the reading by 10.
4. For best accuracy, repeat the complete test with a fresh 
sample.
5. When you’re fi nished testing, remove the Range Extender. 
Thoroughly rinse the cell cup and Range Extender with clean 
water (preferably distilled or deionized) to eliminate dried salt 
build-up. This is extremely important when the instrument will 
be used to test high purity water.
CALIBRATING THE RE-10
1. Calibrate the meter (without RE-10) using the appropriate 
Standard Solution.
2. Fill the cell cup with the appropriate high conductivity 
Standard Solution (442-30,000).
3. Insert RE-10 and press the black button. Multiply the 
reading by 10 and compare it to the value on the Standard 
Solution label. If they are not the same, the RE-10 must be 
recalibrated.
NOTE: DO NOT adjust the Calibration Control.
4. Adjust the white insert of the Extender as follows: If the 
reading is too high—push or tap inward.
If the reading is too low—twist or pull outward with pliers.
NOTE: Each Range Extender is factory calibrated to a particular 
meter. It should be recalibrated if it is to be used with another 
meter.
3






