VII. CALIBRATION
A.
Generally, calibration is recommended about once per month with
B. II
1. Calibration Steps
a. Starting Calibration
Calibration is begun by pressing while measuring Conductivity, TDS
CAL” icon is on, indicating
calibration is now changeable.
The reading is changed with the and keys to match the known value.
The calibration for each of the 4 solution types may be performed in
either conductivity or TDS mode.
the calibration procedures.
KCl, NaCl or 442 User
Cond Gain only Tempco, then Gain
Res Done in conductivity Done in conductivity or TDS
TDS Gain only Tempco, Ratio, then Gain
pH 7, acid and/or base (6Pfc)
ORP Zero set with pH 7 automatically (6Pfc)
Once in CAL mode, the key becomes an “ACCEPT” key. At
each point, pressing accepts the new calibration value and steps
To bypass a calibration step, simply press to accept the present
value as is.
b. Leaving Calibration
Calibration is complete when the “CAL” icon goes out. Pressing any
measurement key cancels changes not yet accepted and exits calibration
mode.
in place of the 3rd buffer.