Owner manual
16
Calibration Limits
There are calibration limits. A nominal “FAC” value is an ideal value
stored by the factory. Attempts to calibrate too far, up or down, from
there will cause the displayed value to be replaced with “FAC
factory calibration for this measurement. The need to calibrate so
far out that “FAC” appears indicates a procedural problem, incorrect
C. Calibration Procedures
1. Conductivity or TDS Calibration
c. Press or
, then
Figure 5
°C
KCl
COND
CAL
press ; “CAL” icon will
appear on the display
d. Press or to
e. Press
solution type is also to be measured, change solution type now
and repeat this procedure.
User Calibration ConductivityTDS
your standard.
c. Press or , then press
three times in TDS. The “CAL” icon will appear on the display.
d. Press or to step the displayed value toward the
standard’s value or hold a key down to scroll rapidly through
the reading.