4. Press
to enter calibration mode. The “CAL”, “BUFFER”
and “7
Displayed value will be the uncalibrated sensor.
NOTES: 7” and “BUFFER”
II will not adjust.
is defective and needs to be replaced
.
or
NOTE:
cause “FAC” to appear. This indicates the need for sensor replacement
The “FAC”
internal electronic calibration is not intended to replace calibration with
Each “FAC” indicates a
to accept the preset factory value, or you may
reduce your variation from factory setting by pressing or
.
6. Press
exit by pressing any measurement key.
b.
IMPORTANT: Always calibrate or verify your Ultrameter II
buffer solution before adjusting the gain with acid or base buffers, i.e.,
point “Gain” calibration and then the opposite for the 3rd point. The
display will verify that a buffer is in the sensor well by displaying either
“Acd” or “bAS”.
CAL”, “BUFFER” and “Acd” or “bAS”