Owner manual


measurement keys will exit without changing the setting. User
mode “Linc” is now complete. The User mode will now use the
calibration gain constant used for the calibration of the
Standard Solution as outlined above.
C. Canceling User mode Calibration “Linc”
The Ultrameter II must be in “User” linked mode in order to cancel the

1. Press “Linked” measurement key ,
or .
Two solution icons will be shown in the left side of display -
User” and another, e.g., “KCl”.
  key until the menu “Linc
3. Press key; the instrument will display both “SEL” and
the “User
4. Press key until “User” is the only solution icon being
displayed.
  key.
6. The User mode calibration “Linc” has now been canceled.
NOTES:
1. To maintain repeatability, use the same standard solutions for
future calibrations.
 II Gain Factor for User mode is
not available when the calibration linc has been established.

         
perform a calibration of the User mode as described in User

be canceled. See above, “Canceling User mode calibration
“Linc””.
3. Once a “Linc” has been established for User mode , the “Linc
will apply to all measurement modes using User solution
