VII. CALIBRATION
A. Calibration Intervals
Generally, calibration is recommended about once per month with
Conductivity or TDS solutions. Calibration with pH solutions should
           

B. II
1. Calibration Steps
a. Starting Calibration
Calibration is begun by pressing while measuring Conductivity,
TDS or pH. Measuring continues, but the CAL icon is on, indicating
calibration is now changeable.
The reading is changed with the and keys to match the
known value. The calibration for each of the 4 solution types may be
performed in either conductivity or TDS mode.

the calibration procedures.
Once in CAL mode, the key becomes an “ACCEPT” key. At
each point, pressing accepts the new calibration value and steps


To bypass a calibration step, simply press to accept the present
value as is.
b. Leaving Calibration
Calibration is complete when the CAL icon goes out. Pressing any
KCl, NaCl or 442 User
Cond Gain only Tempco, then Gain
Res Done in conductivity Done in conductivity or TDS
TDS Gain only Tempco, Ratio, then Gain
pH 7, acid and/or base (6Psi)
ORP Zero set with pH 7 automatically (6Psi)