Manual


6. Press key to accept the setting. Pressing any of the
measurement keys will exit without changing the setting. “User”
mode “Linc” is now complete. The “User” mode will now use the
calibration gain constant used for the calibration of the
Standard Solution as outlined above.
C. Canceling “User” Mode Calibration “Linc”
The Ultrameter II must be in “User” linked mode in order to cancel the

1. Press “Linked” measurement key , or .
Two solution icons will be shown in the left side of display -
User” and another, e.g., “KCl”.
  key until the menu “Linc
  key; the instrument will display both “SEL” and
the “User” Icon.
4. Press key until “User” is the only solution icon being
displayed.
  key.
6. The “User” mode calibration “Linc” has now been canceled.
NOTES:
1. To maintain repeatability, use the same standard solutions for future
calibrations.
   II Gain Factor for “User” mode is not
available when the calibration linc has been established. The other


         
“User” mode Linc should be canceled. See above, “Canceling “User”
mode calibration “Linc””.


