Manual

varies with concentration and temperature in a non-linear fashion. Other
solutions have more variation still. The Ultrameter II uses corrections
that change with concentration and temperature instead of single
average values. See Chart 1.
C.
resulting compensation
How much error results from treating natural water as if it were KCl at
uncompensated conductivity.
Chart 1
0 5 10 15 20 25 30 35 40 45 50 55 60
1.500%
1.600%
1.700%
1.800%
1.900%
2.000%
2.100%
2.200%
2.300%
2.400%
2.500%
KCl % / °C
% / °C
Temperature